Standard Specification and Test Methods for Metallic Bone Staples


Importancia y uso:

A1.3 Significance and Use

A1.3.1 This test method is used to determine the fatigue resistance of metallic bone staples when subjected to repetitive loading for large numbers of cycles. This information may also be useful for comparing the effect of variations in staple material, geometry, surface condition, or placement under certain circumstances.

A1.3.2 It is essential that uniform fatigue practices be established in order that such basic fatigue data be comparable and reproducible and can be correlated among laboratories.

A1.3.3 The results of fatigue tests are suitable for direct application to design only when the service conditions parallel the test conditions exactly. This test method may not be appropriate for all types of bone staple applications. The user is cautioned to consider the appropriateness of the test method in view of the materials being tested and their potential application.

Subcomité:

F04.21

Volúmen:

13.01

Número ICS:

11.040.30 (Surgical instruments and materials)

Palabras clave:

bending test; bone fixation; fatigue test; fixation devices; metallic bone staples; orthopaedic medical devices; pullout test; soft tissue fixation; surgical implants;

$ 1,315

Agregar al carrito

Norma
F564

Versión
24

Estatus
Active

Clasificación
Specification

Fecha aprobación
2024-12-01