Standard Test Method for Finite Element Analysis (FEA) of Metallic Orthopaedic Total Knee Femoral Components under Closing Conditions


Importancia y uso:

3.1 This standard is applicable to the calculation of stresses seen on a knee femoral component when loaded in a manner described in this test method. This method can be used to establish the worst-case size for a particular implant. When stresses calculated using this method were compared to the stresses measured from physical strain gauging techniques performed at one laboratory, the results correlated to within 9 %.

Subcomité:

F04.22

Referida por:

F3495-23, F3210-22E01

Volúmen:

13.02

Número ICS:

11.040.40 (Implants for surgery, prothetics and orthotics)

Palabras clave:

computational simulation; displacement; FEA; finite element analysis; model validation; model verification; orthopaedic implants; strain; stress;

$ 1,090

Agregar al carrito

Norma
F3161

Versión
24

Estatus
Active

Clasificación
Test Method

Fecha aprobación
2024-08-15