Standard Test Method for Coring Testing of Huber Needles


Importancia y uso:

5.1 This test method determines whether Huber needles are designed and manufactured such that they will not produce a core during simulated implantable port access.

5.2 If a needle produces a core during actual use, leaking of the implantable port may occur. Also, the core may be flushed into the port’s reservoir and subsequently into the patient’s body.

Subcomité:

F04.31

Volúmen:

13.02

Número ICS:

11.040.30 (Surgical instruments and materials)

Palabras clave:

core; Huber needle; medical device; non-coring needle;

$ 1,194

Agregar al carrito

Norma
F3212

Versión
16(2023)

Estatus
Active

Clasificación
Test Method

Fecha aprobación
2023-02-01