Standard Guide for Using a Force Tester to Evaluate the Performance of a Brush Part Designed to Clean the External Surface of a Medical Device
Importancia y uso:
5.1 This guide provides two test methods for evaluating the performance characteristics of a brush part designed to clean external surface(s) of a medical device by utilizing force testers.
5.1.1 The first test method utilizes a force tester to measure the force required to actuate a brush part across a surface. This is an indicator of the friction a brush exerts on a surface, a parameter of cleaning effectiveness.
5.1.2 The second test method measures the removal of soil from a surface by a brush part actuated across the surface. This is a further indicator of the effectiveness of a brush part to loosen and remove soil from a surface.
5.2 By providing objective, repeatable methods for evaluating performance under test conditions, this guide can improve the ability to assess the effectiveness of various brush part designs
Subcomité:
F04.15
Referida por:
F3602-23, F3487-20
Volúmen:
13.02
Número ICS:
11.040.01 (Medical equipment in general)
$ 1,194
Norma
F3276
Versión
22
Estatus
Active
Clasificación
Guide
Fecha aprobación
2022-01-01