Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials
Importancia y uso:
4.1 This test method is designed solely for determining comparative laboratory indices of performance. The results may be used for ranking alloys in order of increasing resistance to pitting and crevice corrosion under the specific conditions of this method. It should be noted that the method is intentionally designed to reach conditions that are sufficiently severe to cause breakdown of at least one alloy (Type 316 L stainless steel) currently considered acceptable for surgical implant use, and that those alloys which suffer pitting or crevice corrosion during the more severe portions of the test do not necessarily suffer localized corrosion when placed within the human body as a surgical implant.
Subcomité:
F04.15
Referida por:
F1875-98R22, F2887-23, F2665-21, F1378-18E01, F2229-21, F1586-21, F1314-24, F2581-12R17, G0215-17R24
Volúmen:
13.01
Número ICS:
11.040.40 (Implants for surgery, prothetics and orthotics)
Palabras clave:
crevice corrosion; passive metals; pitting corrosion; potentiostatic polarization measurements; surgical implant materials;
$ 1,090
Norma
F746
Versión
04(2021)
Estatus
Active
Clasificación
Test Method
Fecha aprobación
2021-08-01